Omron Comp Air NE-C801KD - Инструкция по эксплуатации - Страница 21

Небулайзеры Omron Comp Air NE-C801KD - инструкция пользователя по применению, эксплуатации и установке на русском языке. Мы надеемся, она поможет вам решить возникшие у вас вопросы при эксплуатации техники.

Если остались вопросы, задайте их в комментариях после инструкции.

"Загружаем инструкцию", означает, что нужно подождать пока файл загрузится и можно будет его читать онлайн. Некоторые инструкции очень большие и время их появления зависит от вашей скорости интернета.
Страница:
/ 224
Загружаем инструкцию
background image

21

EN

8. Troubleshooting

If any of the below problems occur during use, first check that no other electrical device is 

within 30 cm. If the problem persists, please refer to the table below.

Note:

If the suggested remedy does not solve the problem, do not try to repair the device - no parts of 
the unit are user serviceable. 
Return the unit to an authorized OMRON retail outlet or distributor.

Problem

Cause

Remedy

Nothing happens when the 

power switch is pressed.

Is the AC adapter connected 

correctly to the an electrical 

outlet and the compressor?

Check that the plug is 

inserted in an electrical 

outlet. Unplug then reinsert 

the plug if necessary.

No nebulization or low 

nebulization rate, when the 

power is on.

Is there medication in the 

medication tank?

Add the correct amount of 

medication to the medication 

tank.

Is there too much/little 

medication in the medication 

tank?
Is the vaporizer head missing 

and not assembled correctly?

Attach the vaporizer head 

correctly. 

Is the nebulizer kit 

assembled correctly?

Assemble the nebulizer kit 

correctly.

Is the nozzle blocked?

Make sure that the nozzle is 

free of blockages.

Is the nebulizer kit tilted at a 

sharp angle?

Make sure that the nebulizer 

kit is not tilted at an angle of 

more than 45 degrees.

Is the air tube connected 

correctly?

Make sure that the air tube is 

correctly connected to the 

compressor and nebulizer kit.

Is the air tube folded or 

damaged?

Make sure that the air tube 

does not contain kinks.

Is the air tube blocked?

Make sure that the air tube is 

free of blockages.

Is the air filter dirty?

Replace the air filter with a 

new one.

The device is abnormally 

loud.

Is the air filter cover attached 

correctly?

Attach the air filter cover 

correctly.

The device is very hot.

Is the compressor covered?

Do not cover the compressor 

with any type of cover during 

use.

Is the device operating 

continuously over 20 

minutes?

Limit use to 20 minutes at a 

time, and allow a 40 minute 

interval before using the 

device again.

NE-C801S-KDE(V)_E_M.fm  Page 21  Friday, January 11, 2019  9:55 AM

Содержание

Полезные видео

Характеристики

Оцените статью
tehnopanorama.ru
Остались вопросы?

Не нашли свой ответ в руководстве или возникли другие проблемы? Задайте свой вопрос в форме ниже с подробным описанием вашей ситуации, чтобы другие люди и специалисты смогли дать на него ответ. Если вы знаете как решить проблему другого человека, пожалуйста, подскажите ему :)

Задать вопрос

Часто задаваемые вопросы
Как посмотреть инструкцию к Omron Comp Air NE-C801KD?
Необходимо подождать полной загрузки инструкции в сером окне на данной странице
Руководство на русском языке?
Все наши руководства представлены на русском языке или схематично, поэтому вы без труда сможете разобраться с вашей моделью
Как скопировать текст из PDF?
Чтобы скопировать текст со страницы инструкции воспользуйтесь вкладкой "HTML"